Sometimes, people use D2O as solvent and there is a lot of water in
it. So the solvent peak is so huge and this blocked people from seeing
the interested peaks. There is a way to suppress the solvent peak. The
following procedure assumes that you have already got a standard 1D
spectrum. 1. type dg , enter¿;
This will let you see all the parameters.
2. type dn='H1' enter¿,
place the cursor at the solvent peak and type nl. If you can't
see your spectrum, type ds enter¿;
then place the cursor at the solvent peak; 3. type sd, enter¿;
you will see that dof is changed to a specific number which is exactly
the frequency of the solvent peak. 4. type seqfil='PRESAT' enter¿;;
5. type satmode='y' satpwr=2 satdly=1.5
satfrq= xxx (the number of dof ) or you can
type satfrq=dof,
enter¿;; 6. type dps enter¿;;
you should see that the sequence has a 1.5 second pulse, a soft pulse at
a very low power before the 90 degree pulse. Now you can type dg
to take a look at the parameters. You may change nt, bs. 7. type ga enter¿;;
8. type wft enter¿;;
You should see that your solvent is suppressed. Remember,
interested peaks close to the solvent peak may also be suppressed a
little bit. So integration might be affected. If you still have
problems, come see me. |