
Wecome to our Mass Spectrometer Website
|
Contact manager: 443-854-4640(cell phone) Cungen Zhang, Ph.D czhang17@jhmi.edu
|
The API 150 EX is being checked out everyday by Dr. Cungen Zhang, Manager I. Peptide User: Peptides are usually good at about 50 micrograms/20 microliters -- should get 10^7 cpm comfortably -- but even 10-50x lower concentration should be detectable with MCA signal averaging. The following is the washing procedure: 1. Make sure that your peptide has a best solvent, say Acetone Nitrile. In that case you should wash with 100% CH3-CN first to monitor the decrease of your peptide peaks. You may need to fill in another syringe of pure CH3-CN. Then, wash it with 1:1 of CH3CN:H2O. Then wash it with 100% MeOH and then wash it with 1:1 of MeOH:H2O. 2. If your peptide is most soluble in H2O, then wash it with 100% H2O, and monitor the decrease of the peaks of your peptide. Wash two syringes of the best solvent and then wash with CH3CN, then MeOH and then the mix of MeOH and H2O. 3. Make sure that there is no salt in the solvent. You know Na2CO3 or NaHCO3 can be possibly crystallized in the tubing. So wash completely with H2O if you know that there is some salt in your buffer. 4. Peptide users are strongly suggested to minitor the low molecular region, say 100 to 500. And wash out all the peaks to as low as e^5. e^4 is the ideal level for you to leave.
II. Small Molecule User:
Small organics, the same, in volatile non-flammable solvents (MeOH and MeCN
are ok, though). Sometimes 2-10 x higher concentration for small molecules
is necessary because noise is higher at <1000 amu, however. DMSO/DMF
co-solvents ok at less than 10% v/v.
5. Make sure that your organic coumpound has a best solvent, say CHCl3. In that case you should wash with 100% CHCl3 first to monitor the decrease of your compound peaks. You may need to fill in another syringe of pure CHCl3. (Then, wash it with the second best solvent, say CH3CN and then wash 1:1 of CH3CN:H2O.) Then wash it with 100% MeOH and then wash it with 1:1 of MeOH:H2O.
User Young is working on the spectrometer ----------------------The following are the suggestions by Chris Gross.
When
done, the instrument should be washed with 1:1 MeOH/Water. This wash should
be complete . Usually 3 times. We need to monitor the decreasing of the
peaks.
Other tips:
--------------------------------------------
|
This website was updated on Nov 14, 2006 by Cungen
Zhang. All rights reserved. |